Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754 Recall
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
FDA recall Class II
Ongoing
FDA lists this Class II recall for Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1492-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 3, 2025
- FDA report date
- Apr 9, 2025
- Quantity in commerce
- 37 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
Affected model & UDI checker
Search the exact identifiers published for Z-1492-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754 | Check source notice | 00886799001882 |
FDA code information
UDI/DI 00886799001882, Serial Number range: H0050095-H0050110, H0050112-H0050116, H0050118-H0050124, H0050126, H0050131-H0050134, H0050136, H0050140-H0050142
Why was this product recalled?
During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.
What actions did Terumo Cardiovascular Systems Corporation take?
Terumo Cardiovascular issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 03/03/2025 via email. The notice explained the reason for the removal, potential hazard, and requested the following: 1. Review this Medical Device Correction and assure that all users have received notice of this issue. 2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. 3. Terumo CVS will contact users to discuss removal and correction activities. Note: Terumo CVS recommends that users continue using CDI OneView System H/S Probes while waiting for this removal
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1492-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.