Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1494-2025

Brand Name: ID Circuit Product Name: ID Circuit Flex 220 Recall

Brand Name: ID Circuit Product Name: ID Circuit Flex 220 Model/Catalog Number: MP01347 Software Version: Not applicable Product Description: Breathing circuit for conveying breathing gases between an anesthesia machine or ventilator and adult Component: Not applicable

FDA recall Class I

Ongoing

FDA lists this Class I recall for Brand Name: ID Circuit Product Name: ID Circuit Flex 220. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1494-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Mar 12, 2025
FDA report date
Apr 9, 2025
Quantity in commerce
3325 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: ID Circuit Product Name: ID Circuit Flex 220 Model/Catalog Number: MP01347 Software Version: Not applicable Product Description: Breathing circuit for conveying breathing gases between an anesthesia machine or ventilator and adult Component: Not applicable

Affected model & UDI checker

Search the exact identifiers published for Z-1494-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MP01347Brand Name: ID Circuit Product Name: ID Circuit Flex 220Check source noticeNot specified
Not specifiedBrand Name: ID Circuit Product Name: ID Circuit Flex 220Check source notice04048675249337
FDA code information

Model/Catalog Number: MP01347; UDI Number 04048675249337; All lot numbers.

View the original FDA recall notice

Why was this product recalled?

The potential for cracks forming in the breathing circuit hose.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Draeger, Inc. take?

On March 12, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be Taken When using these breathing circuits: 1. Use holders with a larger contact surface (see Figure 2 above). 2. Perform leak testing after stretching the flexible hose to the required length for the application. Please observe the products Instructions for Use (IFU) According to our records, you have received at least one of the listed breathing circuits. Please inform all potential users in your facility. If you have forwarded affected products to third parties, please also forward this Urgent Medical Device Recall notification. Please also complete the enclosed Medical Device Recall Return Response Acknowledgment and Receipt Form to confirm to us that you have received this information. Adverse events or quality problems experienced with the use of this product may be reported to FDA s MedWatch Adverse Event Reporting program either online or by phone at 1-800-FDA-1088. We apologize for any inconveniences resulting from this action. If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 267-664-1131 or via email at [email protected]. The US Food and Drug Administration has been advised of this action. ****Update 07/08/2025****On June 25, 2025 URGENT MEDICAL DEVICE RECALL - UPDATE letters were sent to consignees to clarify and provide complete UDI information for the device to include both primary and secondary UDI numbers as errors were identified in the initial letter.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology