10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Recall
10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.
FDA recall Class II
Ongoing
FDA lists this Class II recall for 10mL Non-Sterile BD Luer-Lok Tip Syringe Only.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1496-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 27, 2026
- FDA report date
- Mar 18, 2026
- Quantity in commerce
- 204,000 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.
Affected model & UDI checker
Search the exact identifiers published for Z-1496-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 301029 | 10mL Non-Sterile BD Luer-Lok Tip Syringe Only. | Check source notice | Not specified |
FDA code information
Catalog Number: 301029., UDI-DI: N/A. Lot Number: 5268614. Expiration Date: 08/31/2030.
Why was this product recalled?
During the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
What actions did Becton Dickinson & Company take?
Becton Dickinson notified consignees on about 01/27/2026 via emailed URGENT: Medical Device Recall (Correction) letter. Consignees were instructed to visually verify that the correct syringe is selected for the intended clinical application prior to use and that it forms a secure connection, destroy any luer slip syringes found in inventory, circulate the notification to all personnel who need to be aware or to any organization where the potentially affected products have been transferred, and complete and return the provided Customer Response Form. Distributors and kit manufactures were instructed to identify all customers within their distribution network that purchased any affected product and provide a copy of the customer letter to all customers to advise them of this field action notification on BD s behalf. Becton Dickinson is investigating the root cause and will take corrective actions to prevent recurrence of the issue as well as will provide replacement for all discarded inventory.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1496-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology