Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1496-2026

10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Recall

10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.

FDA recall Class II

Ongoing

FDA lists this Class II recall for 10mL Non-Sterile BD Luer-Lok Tip Syringe Only.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1496-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 27, 2026
FDA report date
Mar 18, 2026
Quantity in commerce
204,000 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.

Affected model & UDI checker

Search the exact identifiers published for Z-1496-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
30102910mL Non-Sterile BD Luer-Lok Tip Syringe Only.Check source noticeNot specified
FDA code information

Catalog Number: 301029., UDI-DI: N/A. Lot Number: 5268614. Expiration Date: 08/31/2030.

View the original FDA recall notice

Why was this product recalled?

During the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.

What actions did Becton Dickinson & Company take?

Becton Dickinson notified consignees on about 01/27/2026 via emailed URGENT: Medical Device Recall (Correction) letter. Consignees were instructed to visually verify that the correct syringe is selected for the intended clinical application prior to use and that it forms a secure connection, destroy any luer slip syringes found in inventory, circulate the notification to all personnel who need to be aware or to any organization where the potentially affected products have been transferred, and complete and return the provided Customer Response Form. Distributors and kit manufactures were instructed to identify all customers within their distribution network that purchased any affected product and provide a copy of the customer letter to all customers to advise them of this field action notification on BD s behalf. Becton Dickinson is investigating the root cause and will take corrective actions to prevent recurrence of the issue as well as will provide replacement for all discarded inventory.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.