IVX Station Reconstitution Transfer Set. Recall
IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf boxes per case box. IVX Station Reconstitution Transfer Set is used for the reconstitution of drug vials in hospital pharmacies. It is indicated for use with the IVX Station Pharmacy Compounding System.
FDA recall Class II
Ongoing
FDA lists this Class II recall for IVX Station Reconstitution Transfer Set.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1532-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 20, 2025
- FDA report date
- Apr 16, 2025
- Quantity in commerce
- 3
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf boxes per case box. IVX Station Reconstitution Transfer Set is used for the reconstitution of drug vials in hospital pharmacies. It is indicated for use with the IVX Station Pharmacy Compounding System.
Affected model & UDI checker
Search the exact identifiers published for Z-1532-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 298921215 | IVX Station Reconstitution Transfer Set. | Check source notice | Not specified |
FDA code information
Model number: 298921215. UDI Number: 0086000726042101124301840119041525, 0086000726042101124300840119041525, 0086000726042101124301680119041525. Lot number 011243.
Why was this product recalled?
IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.
What actions did Omnicell, Inc. take?
Single customer was notified on 02/21/2025 via email and later received the letter via hand delivery on 02/25/2025. The notification instructed the customer that Omnicell personnel will remove the affected units, and any Reconstitution Transfer Sets required for Omnicell use during set up and upgrade will be provided as part of a supply kit and will be marked not for human use. The customer was requested to complete and return the Customer Acknowledgement and Receipt.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1532-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology