GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A). Recall

GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).

FDA recall Class II

Ongoing

FDA lists this Class II recall for GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1544-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 20, 2025
FDA report date
Apr 16, 2025
Quantity in commerce
70
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).

Affected model & UDI checker

Search the exact identifiers published for Z-1544-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
K1201-64105GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).Check source noticeNot specified
Not specifiedGenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).Check source notice05700571112601
FDA code information

UDI-DI: 05700571112601. REF/Lot: K1201-64105/0006798023, 5190-7317/0006793917

View the original FDA recall notice

Why was this product recalled?

DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Agilent Technologies, Inc. take?

On 3/20/2025, "URGENT MEDICAL DEVICE RECALL" notices dated 3/20/2025 were emailed to customers who were asked to do the following: 1. Please check your inventory and identify if you have the affected product. 2. Confirm the number of affected kits in your possession and discard any unused kits. 3. Inform those who need to be aware of this notification within your organization or any other organization to which the affected product(s) have been transferred. Please ensure that your organization maintains awareness of this notice. 4. Complete and return the acknowledgement form via email to [email protected] This issue is limited only to chromosome 12, and the data generated can be analyzed using the CytoDx functionalities by displaying the moving average line in the Genome View, Chromosome View, or Gene View of the Triage View. If you have any questions, contact Agilent Technologies, Inc. via email: [email protected] or call 408-386-7294.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology