FDA lists this Class II recall for MEDLINE ReNewal Coblator II ENT Evac 70 Xtra HP,w/Integrated Cable (Blue/Grey), Number REF EICA58740R. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1547-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 5, 2025
FDA report date
Apr 16, 2025
Quantity in commerce
020 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

MEDLINE ReNewal Coblator II ENT Evac 70 Xtra HP,w/Integrated Cable (Blue/Grey), Number REF EICA58740R

Affected model & UDI checker

Search the exact identifiers published for Z-1547-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
EICA58740RMEDLINE ReNewal Coblator II ENT Evac 70 Xtra HP,w/Integrated Cable (Blue/Grey), Number REF EICA58740RCheck source noticeNot specified
Not specifiedMEDLINE ReNewal Coblator II ENT Evac 70 Xtra HP,w/Integrated Cable (Blue/Grey), Number REF EICA58740RCheck source notice10888277395367
FDA code information

UDI/DI 10888277395367, Lot Numbers: 513993, 514445, 514446, 514449, 515558, 515650, 515861, 515902, 515907, 516478, 516549, 517320, 518476, 519258, 519265, 519678, 519759, 519760, 520279, 520282, 520357, 520692, 521356, 521359, 522140, 522141, 522143, 522507, 522508, 524010, 524011, 524015, 524797, 524798, 524802, 528455, 528586, 528587, 528896, 528904, 529932, 530010, 531025, 531027, 532815, 532817.

View the original FDA recall notice

Why was this product recalled?

Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.

What actions did MEDLINE INDUSTRIES, LP - Northfield take?

Medline issued a "Medline ReNewal IMMEDIATE ACTION REQUIRED" notice dated 3/5/2025, to its consignees on 03/05/2025 via USPS first class mail. The notice explained the issue, potential risk, and requested the following actions: REQUIRED ACTION: 1. Immediately check your stock for the affected item numbers and the affected lot numbers. Quarantine all affected product. 2. Please return the completed enclosed response form listing the quantity of affected product on hand. Upon receipt of your submitted response form, FedEx return label(s) will be provided. Even if you do not have any affected product, please complete and return the form, as indicated on the response form. 3. If you have transferred possession of this product to another individual, to another department or another location in your organization you are required to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline ReNewal. For questions, please contact 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology