FDA RECALL RECORD · Z-1551-2025

Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223. Recall

Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1551-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 11, 2025
FDA report date
Apr 16, 2025
Quantity in commerce
N/A
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.

Affected model & UDI checker

Search the exact identifiers published for Z-1551-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedAtellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.Check source notice00630414610887
FDA code information

Siemens Material Number (SMN): 11537223; UDI-DI: 00630414610887; Lot numbers: All lots.

View the original FDA recall notice

Why was this product recalled?

Incorrect software flagging may occur for the assay that may potentially lead to an erroneous result.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Healthcare Diagnostics, Inc. take?

On March 11, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " For both Atellica CH and Atellica CI analyzers, perform the instructions in Appendix B to temporarily reduce the measuring interval. o Until the measuring interval is restored, track additional reagent consumption as a result of these actions to report to Siemens Healthineers for future reimbursement/credit. " Additionally, for Atellica CH Analyzers, perform the instructions in Appendix C to remove rules for flagging of No Calculation results and to install Atellica Solution Software version 1.29.0 or higher. " Complete and return the Field Correction Effectiveness Check form attached to this letter within 30 days. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution A follow-up communication will be provided when Customer Actions are no longer required. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology