Artelon FLEXBAND TWIST .12 Ref: TW012 3.85x17mm HEX Anchor (Qty 2) 5.0x17mm HEX Anchor (Qty 1) 0.5x12cm FLEXBAND (Qty 1) Recall
Artelon FLEXBAND TWIST .12 Ref: TW012 3.85x17mm HEX Anchor (Qty 2) 5.0x17mm HEX Anchor (Qty 1) 0.5x12cm FLEXBAND (Qty 1)
FDA recall Class II
Ongoing
FDA lists this Class II recall for Artelon FLEXBAND TWIST .12 Ref: TW012 3.85x17mm HEX Anchor (Qty 2) 5.0x17mm HEX Anchor (Qty 1) 0.5x12cm FLEXBAND (Qty 1). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1556-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 6, 2026
- FDA report date
- Mar 18, 2026
- Quantity in commerce
- N/A
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Artelon FLEXBAND TWIST .12 Ref: TW012 3.85x17mm HEX Anchor (Qty 2) 5.0x17mm HEX Anchor (Qty 1) 0.5x12cm FLEXBAND (Qty 1)
Affected model & UDI checker
Search the exact identifiers published for Z-1556-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| TW012 | Artelon FLEXBAND TWIST .12 Ref: TW012 3.85x17mm HEX Anchor (Qty 2) 5.0x17mm HEX Anchor (Qty 1) 0.5x12cm FLEXBAND (Qty 1) | Check source notice | Not specified |
| Not specified | Artelon FLEXBAND TWIST .12 Ref: TW012 3.85x17mm HEX Anchor (Qty 2) 5.0x17mm HEX Anchor (Qty 1) 0.5x12cm FLEXBAND (Qty 1) | Check source notice | 00850003396248 |
FDA code information
Lot: RK29976/GTIN: 00850003396248
Why was this product recalled?
Augmentation devices failed bacterial endotoxin testing.
What actions did INTERNATIONAL LIFE SCIENCES take?
On February 6, 2026, International Life Science ( subsidiary of Stryker) issued a Urgent: Medical Device Recall notification to affected consignees. International Life Science asked consignees to take the following actions: 1. Immediately check your internal inventory to locate the product, remove them from their point of use, and isolate/quarantine the unit(s) to prevent accidental use. 2. Review the product issue and risks as communicated and communicate the issue as appropriate in your organization. 3. Return the enclosed business reply form by email a. Response is required, even if you may not have any physical inventory on site anymore. 4. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return of your product(s). 5. Maintain awareness of this communication internally until required actions have been completed within your facility. 6. If you have further distributed the affected product, please notify the applicable parties. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details, so that we can inform the recipients appropriately. b. If you are a distributor, note that you are responsible for notifying your affected customers.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1556-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology