Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1571-2026

Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150 Recall

Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150

FDA recall Class II

Ongoing

FDA lists this Class II recall for Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1571-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 30, 2026
FDA report date
Mar 25, 2026
Quantity in commerce
40 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150

Affected model & UDI checker

Search the exact identifiers published for Z-1571-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
2088026-026Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150Check source noticeNot specified
FDA code information

Product ID Numbers: 4446-2-Centricity Universal Viewer-00558643 100256-3-Centricity Universal Viewer-00558700 101044-1-Centricity Universal Viewer-00558694 13240-2-Centricity Universal Viewer-00558103 4441-1-Centricity Universal Viewer-00558677 4578-1-Centricity Universal Viewer-00558340 10831-2-Centricity Universal Viewer-00558682 656341-Centricity Universal Viewer-01431499 50567-Centricity Universal Viewer-02176369 2690973-Centricity Universal Viewer-01454622 EG1526-Centricity Universal Viewer-00158976 A91901-Centricity Universal Viewer-00044549 M3488569-Centricity Universal Viewer-00071707 25540620 - HC4105-Centricity Universal Viewer-00001393 416005-Centricity Universal Viewer-00260808 1868708X - 556162-Centricity Universal Viewer-00157857 HKOB01-Centricity Universal Viewer-01282268 BMH-Centricity Universal Viewer-01820534 JIPMER-Centricity Universal Viewer-01820555 4347967-Centricity Universal Viewer-01474540 651342-Centricity Universal Viewer-01365218 651342-Centricity Universal Viewer-01365213 A3431341-Centricity Universal Viewer-00545677 473265-Centricity Universal Viewer-01442014 JP-131575-Centricity Universal Viewer-01275307 442900-Centricity Universal Viewer-01460668 577127-Centricity Universal Viewer-01444349 349551-Centricity Universal Viewer-02350518 PL3277-Centricity Universal Viewer-00038831 RU3866-Centricity Universal Viewer-01445731 RU8945-Centricity Universal Viewer-01445740 RU2580-Centricity Universal Viewer-01445743 1354732-Centricity Universal Viewer-01273292 RU3453-Centricity Universal Viewer-01273249 SA2160-Centricity Universal Viewer-00111827 SA1208-Centricity Universal Viewer-00066587 882983-Centricity Universal Viewer-02277629 ZA2533-Centricity Universal Viewer-00116577

View the original FDA recall notice

Why was this product recalled?

There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Medical Systems, LLC take?

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 1/30/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue using your device by following the instructions below: Ensure workstations have appropriate security controls per product manuals. Implement network account authentication by using Active Directory/LDAP services for user management. If network account authentication is not possible at your site, please contact GE HealthCare Service to request temporary steps to mitigate the issue. Ensure all potential users in your facility are informed of this safety notification. Please retain this document for your records. Please complete and return the attached acknowledgement form to [email protected]. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology