ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C Recall
ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C
FDA recall Class I
Ongoing
FDA lists this Class I recall for ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1576-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Feb 13, 2026
- FDA report date
- Apr 1, 2026
- Quantity in commerce
- 131
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C
Affected model & UDI checker
Search the exact identifiers published for Z-1576-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| DC01455550/C | ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C | Check source notice | Not specified |
| DC21452419/C | ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C | Check source notice | Not specified |
| DC21452419-NE5/C | ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C | Check source notice | Not specified |
| DC21452823-NE5/C | ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C | Check source notice | Not specified |
| DC21454035/C | ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C | Check source notice | Not specified |
| DC21455550/C | ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C | Check source notice | Not specified |
FDA code information
REF:UDI-DI/Lot(Expiration): DC01455550/C:00884450697870/H2600169; DC21452419/C:00884450697900/H2659228: DC21452419-NE5/C:00884450697894/H3249278): DC21452823/C:00884450697924/H2600164, H2628890; DC21452823-NE5/C:00884450697917/H2990807: DC21453227/C:00884450697948/H2630602; DC21454035/C:00884450697962/H2628887: DC21455550/C:00884450697979/H2630553, H2659229
Why was this product recalled?
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Merit Medical Systems, Inc. take?
On 2/13/2026, recall notices were mailed and emailed to customers who were asked to do the following: A picture of the 16F Dual-Valved Splittable Sheath Introducer is attached to help you identify the product. Place this notice and the picture on or near the affected product in your inventory (if applicable). Discontinue the use of the 16F Dual-Valved Splittable Sheath Introducer and destroy the sheath introducer at the point of use. Ensure that applicable personnel within your organization are made aware of this field action. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. Complete and return the customer response form via email to [email protected] If you have any questions concerning this communication, contact the firm's Customer Service via email at [email protected] or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri. On 4/2/2026, revised notices were mailed to customers that included stickers to be placed on affected devices to ensure awareness of the recall.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1576-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology