Brand Name: Rover Product Name: Mobile X-ray System Recall
Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Rover Product Name: Mobile X-ray System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1579-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 28, 2025
- FDA report date
- Apr 23, 2025
- Quantity in commerce
- 20 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A
Affected model & UDI checker
Search the exact identifiers published for Z-1579-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MXU-RV35 | Brand Name: Rover Product Name: Mobile X-ray System | Check source notice | Not specified |
FDA code information
Serial Number/UDI: MSN0348 (01)09357123000037(11)230123(21)00348; MSN0349 (01)09357123000037(11)230209(21)00349; MSN0351 (01)09357123000037(11)230413(21)00351; MSN0359 (01)09357123000037(11)231130(21)00359; MSN0361 (01)09357123000037(11)230503(21)00361; MSN0365 (01)09357123000037(11)230510(21)00365; MSN0391 (01)09357123000037(11)230929(21)00391; MSN0406 (01)09357123000037(11)240301(21)00406; MSN0416 (01)09357123000037(11)240429(21)00416; MSN0419 (01)09357123000037(11)240628(21)00419; MSN0428 (01)09357123000037(11)240823(21)00428; MSN0429 (01)09357123000037(11)240823(21)00429; MSN0430 (01)09357123000037(11)240824(21)00430; MSN0431 (01)09357123000037(11)240826(21)00431; MSN0432 (01)09357123000037(11)241108(21)00432; MSN0434 (01)09357123000037(11)241115(21)00434; MSN0435 (01)09357123000037(11)241118(21)00435; MSN0436 (01)09357123000037(11)241122(21)00436; MSN0437 (01)09357123000037(11)241125(21)00437; MSN0438 (01)09357123000037(11)250114(21)00438
Why was this product recalled?
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Micro-X Ltd. take?
On March 28, 2025, Micro-X Ltd. issued a "Urgent: Medical Device Correction" notification to affected consignees via E-Mail. Micro-X asked consignees to take the following actions: 1. Continue to use your product as normal in accordance with its intended use. 2. Your local service representative will be in contact to apply this update to the systems at the next site visit. 3. Circulate this notice to all users of this device so they are aware of the issue. 4. If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. 5. Please retain this letter with your system(s) documentation. 6. Please complete and return the attached response form to Micro-X to the Micro-X LTD in Australia: [email protected] and copy [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1579-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology