Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1585-2025

Brand Name: Novasight Hybrid System Product Name: Novasight Hybrid Catheter Recall

Brand Name: Novasight Hybrid System Product Name: Novasight Hybrid Catheter Model/Catalog Number: TA-06-0001 Product Description: Common Name: Diagnostic intravascular catheter Description: The Novasight Hybrid catheter is a 3Fr single use, sterile device, which will be able to perform transluminal ultrasound and optical coherence tomography imaging of adults. The catheter is capable of real-time 2D side viewing ultrasound and optical imaging acquired simultaneously providing precisely co-registered images with the ability to produce rapidly post processed rendered longitudinal images of the vessel. Packaging: The Novasight Hybrid Catheter is packaged in a packaging system including the protective packaging and the sterile barrier system.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Brand Name: Novasight Hybrid System Product Name: Novasight Hybrid Catheter. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1585-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Mar 12, 2025
FDA report date
Apr 30, 2025
Quantity in commerce
685
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: Novasight Hybrid System Product Name: Novasight Hybrid Catheter Model/Catalog Number: TA-06-0001 Product Description: Common Name: Diagnostic intravascular catheter Description: The Novasight Hybrid catheter is a 3Fr single use, sterile device, which will be able to perform transluminal ultrasound and optical coherence tomography imaging of adults. The catheter is capable of real-time 2D side viewing ultrasound and optical imaging acquired simultaneously providing precisely co-registered images with the ability to produce rapidly post processed rendered longitudinal images of the vessel. Packaging: The Novasight Hybrid Catheter is packaged in a packaging system including the protective packaging and the sterile barrier system.

Affected model & UDI checker

Search the exact identifiers published for Z-1585-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
TA-06-0001Brand Name: Novasight Hybrid System Product Name: Novasight Hybrid CatheterCheck source noticeNot specified
Not specifiedBrand Name: Novasight Hybrid System Product Name: Novasight Hybrid CatheterCheck source notice00628055603054
FDA code information

Model No: TA-06-0001 UDI-DI: (01)00628055603054 Lot number: 230902; Expiration date: 2024-06; Lot number: 240202; Expiration date: 2025-03; Lot number: 240302; Expiration date: 2025-04; Lot number: 240402; Expiration date: 2025-05; Lot number: 240502; Expiration date: 2025-06; and all catheter lots which were expired (See attachment 2a, 2b and 2c for details).

View the original FDA recall notice

Why was this product recalled?

Due to manufacturing issues there is a potential for the catheter sheath to detach.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Conavi Medical, Inc. take?

On 03/12/2025, the firm sent via email an "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" Letter informing customer that the firm has become aware that the sheath of the Novasight Hybrid catheter detached and left in the patient's body during a clinical case but was retrieved during the procedure. Customers are instructed to: Customers shall stop/cease use of the affected the Novasight Hybrid catheters and remove them from service and return immediately to Conavi Medical Inc. This applies to all Novasight Hybrid catheters shipped to U.S.A. regardless of expiry dates. Acknowledge receipt of this recall notification by completing and returning the "Acknowledgement" section of this notification by email to: [email protected]. For questions or assistance - Contact information: Director of Product Management and Marketing Telephone # (416)-483-0100 Ext 109, Email: [email protected] Days/Hours Available for calls: Monday through Friday, 9:00 AM to 4:30 PM Eastern Time.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.