Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1586-2026

Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Recall

Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1586-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 18, 2026
FDA report date
Mar 25, 2026
Quantity in commerce
306
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-1586-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
303585Brand Name: Aptima HPV Assay Product Name: Aptima HPV AssayCheck source noticeNot specified
FDA code information

Lot Code: The Aptima HPV Assay Kit Part Number 303585, Master Lot 929785 contains one Refrigerated Box (PN 302883) and one Room Temperature Box (PN 302885). The following are the UDI number for each box: UDI Number for PN 302883: 15420045500051 UDI Number for PN 302885: 15420045500068

View the original FDA recall notice

Why was this product recalled?

Due to product exhibiting potential to generate either invalid or false negative results.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Hologic, Inc. take?

On 02/18/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter to customers informing them that Hologic has identified that the Aptima HPV Assay kits from ML 929785 exhibit the potential to generate either invalid results for samples and calibrators or false negative results for low analyte samples when reagents are not used on the day of reconstitution. Customers are instructed: 1. Discontinue use of the Aptima HPV Assay kits from ML 929785. 2. Segregate and count the number of affected product on-hand and dispose according t the local guidelines and regulations. 3.Fill out the Customer Response Form using the link https://iqvia-response.my.site.com/mt/fca?cid=hpv-929785 4. Hologic will process and ship the replacement kits For Questions or concerns, contact Hologic Technical Support at +1 888-484-4747 or +1 858-410-8511 during hours 5 am to 5 pm PT or by email at [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology