FDA RECALL RECORD · Z-1610-2026

Olympus Endoscope Suction Valve - MAJ-1443. Recall

Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Olympus Endoscope Suction Valve - MAJ-1443.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1610-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 12, 2026
FDA report date
Apr 1, 2026
Quantity in commerce
47,383 units
Distribution
See key details
FDA cause
Reprocessing Controls

What is being recalled?

Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.

Affected model & UDI checker

Search the exact identifiers published for Z-1610-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MAJ-1443Olympus Endoscope Suction Valve - MAJ-1443.Check source noticeNot specified
N5363430Olympus Endoscope Suction Valve - MAJ-1443.Check source noticeNot specified
Not specifiedOlympus Endoscope Suction Valve - MAJ-1443.Check source notice04953170355912
FDA code information

Model Number: MAJ-1443. UDI-DI: 04953170355912. Catalog Number: N5363430. All Lot/Serial Numbers.

View the original FDA recall notice

Why was this product recalled?

The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees on about 02/12/2026 via URGENT: MEDICAL DEVICE CORRECTION letter. Consignees were instructed to immediately cease reprocessing the MAJ-1443 and MAJ-1444 valves in an OER-Pro or OER-Elite and ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the full extent of the MAJ-1443 and MAJ-1444 reprocessing steps as detailed in the provided Instructions for Use (IFU). The updated IFU can also be accessed electronically through OlympusConnect.com. Additionally, consignees were instructed to replace any existing copies of the MAJ-1443/MAJ-1444 IFU with the updated IFU provided in this letter, identify and notify all customers if affected units were further distributed and complete and confirm receipt of the notification via the Olympus web portal.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology