FDA RECALL RECORD · Z-1618-2026

Philips 3D6-2 Ultrasound Transducer Recall

Philips 3D6-2 Ultrasound Transducer

FDA recall Class III

Ongoing

FDA lists this Class III recall for Philips 3D6-2 Ultrasound Transducer. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1618-2026
Recall classification
Class III
Enforcement status
Ongoing
Initiation date
Sep 5, 2025
FDA report date
Apr 1, 2026
Quantity in commerce
17 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Philips 3D6-2 Ultrasound Transducer

Affected model & UDI checker

Search the exact identifiers published for Z-1618-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
989605326521Philips 3D6-2 Ultrasound TransducerCheck source noticeNot specified
Not specifiedPhilips 3D6-2 Ultrasound TransducerCheck source notice00884838067714
FDA code information

Model No. 989605326521, 989605440872, 989605440871; UDI: (01)00884838067714(21)02Q19C, (01)00884838067714(21)02VF0G, (01)00884838067714(21)02DNKX, (01)00884838067714(21)02BXT7, (01)00884838067714(21)02TWD3, (01)00884838067714(21)02RRQX, (01)00884838067714(21)02HD9F; Serial No. 02V8MV, 02Q19C, 02DNGH, 02TWL6, 02P2LF, 031XK6, 02VF0G, 02HJ9Q, 02DNKX, 02BXT7, 02VHQZ, 02TWD3, 02NDCL, 02CBD4, 02RRQX, 02KWQ9, 02HD9F.

View the original FDA recall notice

Why was this product recalled?

To provide clarification and labelling to define the useful life of ultrasound transducers in the field.

What actions did Philips Ultrasound, Inc take?

An URGENT Medical Device Correction Notice, dated 8/19/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology