Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Recall

Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number: ES-100V3 (device); BT2M20S8C (2 MHz fetal probe). Software Version: N/A Product Description: Hand-held, battery-powered ultrasonic Doppler blood-flow detector with LCD numeric display, used with interchangeable probes to detect vascular blood-flow signals and display heart rate; affected units were packaged with a 2 MHz BT2M20S8C fetal probe and labeled for both vascular and fetal applications.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1631-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 4, 2026
FDA report date
Apr 1, 2026
Quantity in commerce
34
Distribution
See key details
FDA cause
Labeling Change Control

What is being recalled?

Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number: ES-100V3 (device); BT2M20S8C (2 MHz fetal probe). Software Version: N/A Product Description: Hand-held, battery-powered ultrasonic Doppler blood-flow detector with LCD numeric display, used with interchangeable probes to detect vascular blood-flow signals and display heart rate; affected units were packaged with a 2 MHz BT2M20S8C fetal probe and labeled for both vascular and fetal applications.

Affected model & UDI checker

Search the exact identifiers published for Z-1631-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
ES-100V3Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probeCheck source noticeNot specified
FDA code information

Lot Code: BIDOP 3 #00894912002043 BT2M20S8C #00894912002326

View the original FDA recall notice

Why was this product recalled?

Product labeling includes a fetal indication for use that is not cleared under its 510(k).

What actions did Koven Technology, Inc. take?

On March 4, 2026, firm began notifying customers via first-class mail. Customers were informed they may still use the BiDop 3 device for vascular applications, but that the device is not cleared for fetal usage. Koven Technology, Inc. will provide a replacement operation manual for devices that are being amended to remove fetal use indications.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology