Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080 Recall
Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080
FDA recall Class II
Ongoing
FDA lists this Class II recall for Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1644-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 17, 2025
- FDA report date
- Apr 30, 2025
- Quantity in commerce
- 2683
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080
Affected model & UDI checker
Search the exact identifiers published for Z-1644-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 9-TVLPC4F90/080 | Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080 | Check source notice | Not specified |
| Not specified | Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080 | Check source notice | 00811806012926 |
FDA code information
UDI-DI 00811806012926 Lots 9078966 9093895 9134623 9144113 9148110 9148579 9155868 9158302 9241591 9241593 10004059 10011461 10011462 10011463 10011465 10017790 10017792 10196106 10196119 10198795 10198801 10198889 10207472 10207503 10274454 10283818
Why was this product recalled?
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Abbott take?
On March 17, 2025, the firm notified customers via hand-delivered customer communications titled "Urgent Medical Device Recall." Customers were instructed to share the notice with applicable personnel and to return any remaining unused devices from impacted lots. Abbott representatives will assist with product return and replenishment. If you have any questions, please contact your local Sales Representative or Abbott Support at 1-800-544-1664.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1644-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology