Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1644-2025

Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080 Recall

Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080

FDA recall Class II

Ongoing

FDA lists this Class II recall for Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1644-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 17, 2025
FDA report date
Apr 30, 2025
Quantity in commerce
2683
Distribution
See key details
FDA cause
Process control

What is being recalled?

Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080

Affected model & UDI checker

Search the exact identifiers published for Z-1644-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
9-TVLPC4F90/080Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080Check source noticeNot specified
Not specifiedAmplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080Check source notice00811806012926
FDA code information

UDI-DI 00811806012926 Lots 9078966 9093895 9134623 9144113 9148110 9148579 9155868 9158302 9241591 9241593 10004059 10011461 10011462 10011463 10011465 10017790 10017792 10196106 10196119 10198795 10198801 10198889 10207472 10207503 10274454 10283818

View the original FDA recall notice

Why was this product recalled?

Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Abbott take?

On March 17, 2025, the firm notified customers via hand-delivered customer communications titled "Urgent Medical Device Recall." Customers were instructed to share the notice with applicable personnel and to return any remaining unused devices from impacted lots. Abbott representatives will assist with product return and replenishment. If you have any questions, please contact your local Sales Representative or Abbott Support at 1-800-544-1664.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology