Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented Recall
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
FDA recall Class II
Ongoing
FDA lists this Class II recall for Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1645-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 24, 2025
- FDA report date
- Apr 30, 2025
- Quantity in commerce
- 416 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
Affected model & UDI checker
Search the exact identifiers published for Z-1645-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 802602202 | Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented | Check source notice | Not specified |
| Not specified | Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented | Check source notice | 00887868376320 |
FDA code information
GTIN 00887868376320, All lots distributed with IFU 87-6204-051-99 Rev D or prior
Why was this product recalled?
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
What actions did Zimmer, Inc. take?
Zimmer Biomet issued an URGENT MEDICAL DEVICE RECALL CORRECTION - PHASE 1 notice to its consignees on 03/24/2025 via email and courier. The notice explains the issue, potential risk, and requested the following: Distributors: Your Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 3. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. 4. Retain a copy of your Inventory Return Certification and product return forms for your records in the event of a compliance audit of your facility. Risk Manager Responsibilities: 1. Review this correction notice and ensure that affected personnel, including surgeons, are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with this notice ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgeme
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1645-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026