ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700. Recall
ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
FDA recall Class II
Ongoing
FDA lists this Class II recall for ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1653-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 10, 2026
- FDA report date
- Apr 15, 2026
- Quantity in commerce
- U.S. 758, OUS 2077
- Distribution
- See key details
- FDA cause
- Radiation Control for Health and Safety Act
What is being recalled?
ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
Affected model & UDI checker
Search the exact identifiers published for Z-1653-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 10849000 | ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700. | Check source notice | Not specified |
| Not specified | ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700. | Check source notice | 4056869046877 |
| Not specified | ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700. | Check source notice | 4056869063317 |
| Not specified | ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700. | Check source notice | 4056869295923 |
| Not specified | ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700. | Check source notice | 4056869149325 |
FDA code information
ARTIS pheno UDI: 4056869046877. ARTIS icono biplane UDI: 4056869063317. ARTIS icono ceiling UDI: 4056869295923. ARTIS icono floor UDI: 4056869149325.
Why was this product recalled?
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
What actions did Siemens Medical Solutions USA, Inc. take?
Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice, dated March 9, 2026, to all affected customers.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1653-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology