Brand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise Catheter Recall

Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise Catheter. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1655-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 29, 2026
FDA report date
Apr 8, 2026
Quantity in commerce
24
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation

Affected model & UDI checker

Search the exact identifiers published for Z-1655-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PRDS-068-02Brand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise CatheterCheck source noticeNot specified
Not specifiedBrand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise CatheterCheck source notice00810008950043
FDA code information

Lot Code: Model No PRDS-068-02; UDI-DI 00810008950043; Lot Number: M4862

View the original FDA recall notice

Why was this product recalled?

Due to firm distributing a nonconforming product that supposed to be scrapped and not for human use.

What actions did Recor Medical, Inc. take?

On 01/28/2026, the firm informed customers verbally and sent via mail an "MEDICAL DEVICE RECALL" Letter that nonconforming products (cracking in the hub adjacent to the guidewire port) that may lead to fluid leaks and could trigger a pressure error. Customers are instructed to: 1. Do not use the affected catheters for any patient procedures. 2. Identify the catheters from the affected lot and return to Recor Clinical Specialist through RMA process. 3. Review and complete the Acknowledgement and Receipt Form and provide to the Recor Clinical Specialist. For questions or further assistance - contact Director of Regulatory & Quality Compliance at [email protected] or telephone at 650-542-7700

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.