Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1657-2025

MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S; Recall

MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;

FDA recall Class II

Ongoing

FDA lists this Class II recall for MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1657-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 3, 2025
FDA report date
Apr 30, 2025
Quantity in commerce
1 unit
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;

Affected model & UDI checker

Search the exact identifiers published for Z-1657-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MC2-5090SMAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;Check source noticeNot specified
FDA code information

Version/Model: MC2-5090S; UDI-DI (with lot nuber): (01)00887517834041(11)250311(17)300311(10)GB1607ET

View the original FDA recall notice

Why was this product recalled?

Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

What actions did GLOBUS MEDICAL, INC. take?

On April 3, 2025, URGENT: MEDICAL DEVICE NOTIFICATION letter was sent to their customer. Actions to be taken by Customers/Users The customer is instructed to monitor their patient closely for any potential adverse events. Report any adverse events believed to be associated with this notification to Globus Medical. In addition, any adverse events associated with the use of this product and/or quality problems should also be reported to the FDA's MedWatch Program: MedWatch: The FDA Safety Information and Adverse Event Reporting Program; For Questions and Support: Globus Medical Complaint Handling Unit: Email: [email protected] Globus Medical is committed to serving your product needs and our primary objectives are patient safety, exceptional product reliability, and the highest level of customer satisfaction. Thank you for your prompt support on this important matter. We appreciate your cooperation.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology