Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1660-2026

ALLURA Xper FD10F; System Code: 722002; Recall

ALLURA Xper FD10F; System Code: 722002;

FDA recall Class II

Ongoing

FDA lists this Class II recall for ALLURA Xper FD10F; System Code: 722002;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1660-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 3, 2026
FDA report date
Apr 8, 2026
Quantity in commerce
10 units (8 US, 2 OUS)
Distribution
See key details
FDA cause
Device Design

What is being recalled?

ALLURA Xper FD10F; System Code: 722002;

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

System Code: 722002; UDI: N/A; System Serial Number: 53, 61, 78, 38, 28, 63, 3, 81, 66, 40;

View the original FDA recall notice

Why was this product recalled?

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?

On March 3, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue and follow the instructions below. 2. Perform the daily check before starting the first procedure, as described in the device labelling 3. Should you experience a foot switch activation issue, please report it to your local Philips representative. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 5. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 6. Complete and return the response form included in this Urgent Medical Device Correction letter to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology