The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished). Recall
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
FDA recall Class II
Ongoing
FDA lists this Class II recall for The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1665-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 20, 2025
- FDA report date
- Apr 30, 2025
- Quantity in commerce
- N/A
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
Affected model & UDI checker
Search the exact identifiers published for Z-1665-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished). | Check source notice | 00884999047389 |
| Not specified | The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished). | Check source notice | 00884999048034 |
| Not specified | The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished). | Check source notice | 00884999047587 |
FDA code information
All Serial Numbers. 08N53-001, UDI-DI: 00884999047389; 08N53-002, UDI-DI: 00884999048034; 08N53-032, UDI-DI: 00884999047587
Why was this product recalled?
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
What actions did Abbott Molecular, Inc. take?
An Urgent Field Safety Notice/ Field Correction Recall notification letter was sent to customers. Necessary Actions In accordance with the Alinity m Operations Manual, users should always utilize Personal Protective Equipment when engaging with the instrument. In the event you observe a leak, please take appropriate precautions to prevent exposure and immediately contact your Abbott Molecular Representative for additional troubleshooting information. Please complete and return the associated Customer Reply Form. If you have forwarded this product to any other laboratories, please also forward this letter and customer reply form to that laboratory. Review this information with laboratory personnel and retain this communication for future reference. If you have any questions regarding this communication, please contact your local Abbott representative. We apologize for any inconvenience this may have created for your laboratory.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1665-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology