IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application. Recall
IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
FDA recall Class II
Ongoing
FDA lists this Class II recall for IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1689-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 24, 2025
- FDA report date
- May 7, 2025
- Quantity in commerce
- 67 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
Affected model & UDI checker
Search the exact identifiers published for Z-1689-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| H7493967534170 | IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application. | Check source notice | Not specified |
| Not specified | IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application. | Check source notice | 00191506032708 |
FDA code information
GTIN00191506032708, Batch Numbers: 33988638, 33988639, 34309895, 34309896, 34392269, 34392270, 34392271, 34564920.
Why was this product recalled?
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
What actions did BOSTON SCIENTIFIC CORPORATION take?
Boston Scientific issued a Product Advisory notice to its consignees on 03/24/2025 via Fed Ex, email, and hand delivery. The notice explained the issue, clinical impact, and provided the following recommendations: " Avoid disconnecting and reconnecting these IceSeed CX Cryoablation Needles from/to the Cryoablation System. " Maintain initial connection of IceSeed CX Cryoablation Needles to the Cryoablation System and the IceSeed CX Cryoablation Needles will perform as intended. The consignee was also directed to immediately post this information in a visible location near the affected products to ensure it is readily accessible to all handlers and users of the device. Share this notice with any health care professional within your organization and with any entities where the affected devices may have been transferred and complete the acknowledgement form and return it to [email protected] (email) or Fax to: Field Action Center 1-763-415-7708.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1689-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology