Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Recall
Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min) Contains AH165 Circuit, AH290 Chamber Component: No. a conduit for respiratory gas between the patient and a ventilator.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1698-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Apr 10, 2025
- FDA report date
- May 21, 2025
- Quantity in commerce
- 3053
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min) Contains AH165 Circuit, AH290 Chamber Component: No. a conduit for respiratory gas between the patient and a ventilator.
Affected model & UDI checker
Search the exact identifiers published for Z-1698-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| AH265 | Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit | Check source notice | Not specified |
FDA code information
Lot Code: Lot/Serial Number(s): 0004247924 0004247925 0004252940 0004252941 0004256292 0004256293 0004256294 0004256295 0004272325 0004272868 0004278978 0004279156 0004280036 0004285304 0004288036 0004289514 UDI-DI: AirLife Label Each: 10889483595909 Case: 30889483595903 Vyaire Label: Each: 10190752145160 Case: 50190752145168
Why was this product recalled?
Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Vyaire Medical take?
On April 10, 2025, the firm notified customers via emailed letters titled URGENT: MEDICAL DEVICE RECALL. Customers are instructed to examine their inventory for the affected lots of product. Product in inventory should be quarantined. For in-use product - 1) For affected product in use that is utilizing the adaptor connection, immediately stop use of the product. 2) For affected product in use that is not utilizing the adaptor connection, ventilator circuit may continue to be utilized. Customers were provided with additional precautions to take to ensure proper function. Customers can choose between returning or destroying affected product. Firm will provide replacement product. If you have any questions regarding this field action, please call AirLife at 1-800-433-2797, or e-mail at [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1698-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology