Welch Allyn Life2000 Ventilation system, Product Codes: 1) BT-20-0002; 2) BT-20-0002A; 3) BT-20-0002AP; 4) BT200007; 5) BT-20-0007; 6) BT200011; 7) BT-20-0011; 8) RMS010118CP. Recall
Welch Allyn Life2000 Ventilation system, Product Codes: 1) BT-20-0002; 2) BT-20-0002A; 3) BT-20-0002AP; 4) BT200007; 5) BT-20-0007; 6) BT200011; 7) BT-20-0011; 8) RMS010118CP.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Welch Allyn Life2000 Ventilation system, Product Codes: 1) BT-20-0002; 2) BT-20-0002A; 3) BT-20-0002AP; 4) BT200007; 5) BT-20-0007; 6) BT200011; 7) BT-20-0011; 8) RMS010118CP.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1725-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Apr 7, 2025
- FDA report date
- May 21, 2025
- Quantity in commerce
- 4881 units
- Distribution
- See key details
- FDA cause
- Software in the Use Environment
What is being recalled?
Welch Allyn Life2000 Ventilation system, Product Codes: 1) BT-20-0002; 2) BT-20-0002A; 3) BT-20-0002AP; 4) BT200007; 5) BT-20-0007; 6) BT200011; 7) BT-20-0011; 8) RMS010118CP.
Affected model & UDI checker
Search the exact identifiers published for Z-1725-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Welch Allyn Life2000 Ventilation system, Product Codes: 1) BT-20-0002; 2) BT-20-0002A; 3) BT-20-0002AP; 4) BT200007; 5) BT-20-0007; 6) BT200011; 7) BT-20-0011; 8) RMS010118CP. | Check source notice | 00815410020537 |
| Not specified | Welch Allyn Life2000 Ventilation system, Product Codes: 1) BT-20-0002; 2) BT-20-0002A; 3) BT-20-0002AP; 4) BT200007; 5) BT-20-0007; 6) BT200011; 7) BT-20-0011; 8) RMS010118CP. | Check source notice | 00815410020278 |
FDA code information
ALL SERIAL NUMBERS: 1) BT-20-0002, UDI/DI 00815410020537 or 00887761978089; 2) BT-20-0002A, UDI/DI 00815410020537 of 00887761978089; 3) BT-20-0002AP, UDI/DI 00815410020537 or 00887761978089; 4) BT200007, UDI/DI 00815410020278 or 00887761978089; 5) BT-20-0007, UDI/DI 00815410020278 or 00887761978089; 6) BT200011, UDI/DI 00815410020537 or 00887761978089; 7) BT-20-0011, UDI/DI 00815410020537 or 00887761978089; 8) RMS010118CP, UDI/DI 00815410020537 or 00887761978089.
Why was this product recalled?
A cybersecurity vulnerability was discovered through internal testing.
What actions did Baxter Healthcare Corporation take?
Baxter issued an URGENT MEDICAL DEVICE RECALL notice to healthcare providers and patients beginning on 04/07/2025 via USPS first class mail. The notice explained that Baxter was voluntarily initiating a permanent removal of all Life 2000 ventilators currently in use due to a cybersecurity issue and requested the following: Actions to be Taken by DME Providers: 1. Notify your patients using Life2000 ventilators of this recall. 2. Work with patients and prescribers to identify an alternate therapy option. 3. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors. Actions to be Taken by Healthcare Providers: 1. Locate and discontinue use of all Life2000 ventilators and compressors within your facility. 2. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors at your facility. Actions to be Taken by Distributors: 1. Please disseminate this information and conduct a user-level recall of the affected product that you distributed to customers. 2. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors. If you suspect that a device has been exposed to unauthorized personnel, contact Baxter Acute Care Customer Service team. The team can help verify whether the device settings are correct and if the device is performing as expected. The team can be reached by email at [email protected], or by phone at 800-426-4224, option 2, then option 1. For questions regarding this communication, contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, between the hours of 8:00 am and 6:00 pm Central Time, Monday through Friday.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1725-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.