Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1727-2025

iGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle. Recall

iGo¿2 DC Car Adapter to be used with iGo¿2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.

FDA recall Class I

Ongoing

FDA lists this Class I recall for iGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1727-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Apr 11, 2025
FDA report date
May 21, 2025
Quantity in commerce
89,532 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

iGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.

Affected model & UDI checker

Search the exact identifiers published for Z-1727-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
DV6X-619iGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.Check source noticeNot specified
125DiGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.Check source noticeNot specified
Not specifiediGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.Check source notice885304020585
Not specifiediGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.Check source notice885304022237
Not specifiediGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.Check source notice885304032250
Not specifiediGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.Check source notice885304033257
Not specifiediGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.Check source notice885304033240
Not specifiediGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.Check source notice885304033516
FDA code information

DC Cord Code Info: Catalog No. DV6X-619; UDI 885304020585; Cord Revision: Rev E; Lot No. beginning with 2023-01-01 through 2024-12-31. iGo2 Portable Oxygen Concentrator Systems Code Info: Catalog No. 125D, 125D-XB, 125D-BT, 125D-BT-XB, 125D-ARYA-XB; UDI 885304022237, 885304032250, 885304033257, 885304033240, 885304033516; Lot No. D23125001DS to D23A26121DS, D23706001DA to D23B16056DA, F23B06001DS to F25108143DS, F23313001DH to F25109259DH, F24212001DA to F25217150DA.

View the original FDA recall notice

Why was this product recalled?

Potential for DC Power Supply housing to become hot to the touch and deform the plastic housing.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medical Depot Inc. take?

An URGENT NOTICE MEDICAL DEVICE RECALL dated 4/11/25 was emailed to consignees. The recall notification instructs consignees on how to identify affected devices in their possession, destroy the affected charger cord, and provides instructions on how to obtain a replacement charger cord through Drive Devilbiss. If product was further distributed, consignees are to contact their customers; Devilbiss has a draft recall notification and recall website available for their consignees who further distributed product. For those units that are rentals, Devilbiss recommends that consignees check devices as they return from rental use. Dan Hinkbein is available to assist consignees who distributed affected devices fulfill recall actions and can be contacted at [email protected]. Consignees with any issues or questions can call Drive Devilbiss Healthcare at 888-770-7017 Monday through Friday from 7:30 AM to 5:00 PM CST. Questions can also be emailed to [email protected]. The firm has set up a website for this recall event at www.recallrtr.com/iGo2-car-cord.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology