Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1728-2025

Artix MT Thrombectomy Device, REF: 32-102 Recall

Artix MT Thrombectomy Device, REF: 32-102

FDA recall Class II

Ongoing

FDA lists this Class II recall for Artix MT Thrombectomy Device, REF: 32-102. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1728-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 8, 2025
FDA report date
May 14, 2025
Quantity in commerce
45
Distribution
See key details
FDA cause
Labeling Change Control

What is being recalled?

Artix MT Thrombectomy Device, REF: 32-102

Affected model & UDI checker

Search the exact identifiers published for Z-1728-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
32-102Artix MT Thrombectomy Device, REF: 32-102Check source noticeNot specified
Not specifiedArtix MT Thrombectomy Device, REF: 32-102Check source notice00810123710584
FDA code information

UDI-DI: 00810123710584, Lot: 24100082, Expiration: 14 April 2025. IFU: IU-01087 Rev. B

View the original FDA recall notice

Why was this product recalled?

Thrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Inari Medical - Oak Canyon take?

On 4/8/2025, correction and removal notices were distributed to customers who were asked to do the following: 1) Review notice and the accompanying copy of the device IFU (IU-01087 Rev. C) and disseminate it to any applicable personnel and device users at your facility and network to ensure they are aware of the unused device replacement. 2) Notice should be shared with any organization where affected devices may have been transferred. 3) Complete and return the acknowledgement reply form via email to If you have any questions, please contact your local sales representative, or the firm's Customer Service at 877-923-4747, or Quality department at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology