eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152 Recall
eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152
FDA recall Class II
Ongoing
FDA lists this Class II recall for eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1734-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 2, 2026
- FDA report date
- Apr 8, 2026
- Quantity in commerce
- 20 units
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152
Affected model & UDI checker
Search the exact identifiers published for Z-1734-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152 | Check source notice | 00841436111065 |
FDA code information
UDI-DI: 00841436111065; Lot Number: A233347, A228677, A268116, A268238, A272513, A279457, A280388 , A281840
Why was this product recalled?
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did CIVCO Medical Instruments Co., Inc. take?
An URGENT: MEDICAL DEVICE RECALL notification letter dated 3/2/26 was sent to customers. Actions to be taken by the Customer/User: CIVCO requests that you check your inventory for any eTRAX Needle Sensor for Aurora Trackers and complete the supplied response letter for the appropriate action. Please contact your distribution partner to return any existing inventory, or report its destruction by completing the attached response form. Please complete the enclosed response form even if you do not have any eTRAX Needle Sensor for Aurora Trackers remaining in your inventory to assist in our reconciliation process. Response forms can also be sent via email to [email protected]. For questions, contact CIVCO at # 800-445-6741, or 319-248-6757. Type of Action by the Company: This notice is being communicated to all customers who have purchased an eTRAX Needle Sensor for Aurora Trackers from affected lots. CIVCO requests you provide this notification to the appropriate personnel within your facility. If the affected product was distributed outside of your organization, please notify those locations down to the medical facility level.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1734-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology