Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1734-2026

eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152 Recall

eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152

FDA recall Class II

Ongoing

FDA lists this Class II recall for eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1734-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 2, 2026
FDA report date
Apr 8, 2026
Quantity in commerce
20 units
Distribution
See key details
FDA cause
Software change control

What is being recalled?

eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152

Affected model & UDI checker

Search the exact identifiers published for Z-1734-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedeTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152Check source notice00841436111065
FDA code information

UDI-DI: 00841436111065; Lot Number: A233347, A228677, A268116, A268238, A272513, A279457, A280388 , A281840

View the original FDA recall notice

Why was this product recalled?

There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did CIVCO Medical Instruments Co., Inc. take?

An URGENT: MEDICAL DEVICE RECALL notification letter dated 3/2/26 was sent to customers. Actions to be taken by the Customer/User: CIVCO requests that you check your inventory for any eTRAX Needle Sensor for Aurora Trackers and complete the supplied response letter for the appropriate action. Please contact your distribution partner to return any existing inventory, or report its destruction by completing the attached response form. Please complete the enclosed response form even if you do not have any eTRAX Needle Sensor for Aurora Trackers remaining in your inventory to assist in our reconciliation process. Response forms can also be sent via email to [email protected]. For questions, contact CIVCO at # 800-445-6741, or 319-248-6757. Type of Action by the Company: This notice is being communicated to all customers who have purchased an eTRAX Needle Sensor for Aurora Trackers from affected lots. CIVCO requests you provide this notification to the appropriate personnel within your facility. If the affected product was distributed outside of your organization, please notify those locations down to the medical facility level.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.