Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1738-2025

Blueprint Software, BPUE001. Intended for use by surgeons as a pre surgical planner for shoulder replacement. Recall

Blueprint Software, BPUE001. Intended for use by surgeons as a pre surgical planner for shoulder replacement.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Blueprint Software, BPUE001. Intended for use by surgeons as a pre surgical planner for shoulder replacement.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1738-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 10, 2025
FDA report date
May 14, 2025
Quantity in commerce
438 users
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Blueprint Software, BPUE001. Intended for use by surgeons as a pre surgical planner for shoulder replacement.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Software Version 4.2.1

View the original FDA recall notice

Why was this product recalled?

A software bug in Blueprint version 4.2.1 prevents users from accessing cases initiated in previous versions of the software.

What actions did Tornier S.A.S. take?

On April 10, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were informed that an Online Management System (OMS) software patch was released to address the bug on March 25, 2025. The issue can no longer occur with the introduction of this patch. No actions are required on the customer's part to implement the patch. Version 4.2.2 is currently in development and is scheduled for release by end of Quarter 2, 2025. This patch will implement a permanent fix for the issue. Customers will be prompted to complete this update within the Blueprint Software when it is available.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.