Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit. Recall

Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1743-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 31, 2025
FDA report date
May 21, 2025
Quantity in commerce
72 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit.

Affected model & UDI checker

Search the exact identifiers published for Z-1743-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
RPGH9472-01Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit.Check source noticeNot specified
Not specifiedHalyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit.Check source notice10809160211311
FDA code information

Model No. RPGH9472-01; UDI: 10809160211311; Lot Code: Lot: 1636640; Exp: 1/13/2028.

View the original FDA recall notice

Why was this product recalled?

An issue with a raw material used to coat some surgical needles that may result in diminished performance and/or removal of some silicone needle coating during use.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AVID Medical, Inc. take?

An URGENT: MEDICAL DEVICE RECALL NOTIFICATION dated 3/31/25 was emailed to consignees. The notification indicates that this recall is due to Ethicon having recalled PDS II, PDS Plus, PROLENE, and ATHIBOND EXCEL sutures, which are components in the convenience kits being recalled by the recalling firm. Consignees are being instructed to discontinue use of the Johnson & Johnson Ethicon Sutures contained inside the recalled AVID kits identified in the recall notification and to add the provided labels to affected kits to ensure recalled sutures are discarded. Distributors are asked to forward the provided recall notification to their customers to which product was further distributed. Consignees are asked to return the provided response form via email. Questions regarding this recall event are to be directed to your distributor or sales representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology