FDA RECALL RECORD · Z-1747-2025

FoundationOne Companion Diagnostic (F1CDx) Recall

FoundationOne Companion Diagnostic (F1CDx)

FDA recall Class II

Ongoing

FDA lists this Class II recall for FoundationOne Companion Diagnostic (F1CDx). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1747-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 9, 2025
FDA report date
May 21, 2025
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

FoundationOne Companion Diagnostic (F1CDx)

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

ORD-XXXXX28-01 ORD-XXXXX98-01 ORD-XXXXX29-01 ORD-XXXXX99-01 Note: The above characters have been redacted to minimize patient information disclosure.

View the original FDA recall notice

Why was this product recalled?

Four (4) total reports were sent to customers with clinically relevant variants requiring amendment, one (1) of which with an incorrect negative CDx claim. The remaining three (3) with tumor profiling result changes did not impact CDx indications.

What actions did Foundation Medicine, Inc. take?

On April 9, 2025 URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters were sent to customers. Asked Customer Immediate Actions: Foundation Medicine respectfully requests your support with the following actions: 1. Review, complete, sign and return the attached business reply form to be returned in accordance with the directions on the business reply form within 5 business days of receipt of this notification. 2. Forward this notice to anyone in your facility that needs to be informed. 3. If there is any impact that requires this notice to be forwarded to another facility, contact that facility and provide them with this letter. 4. Keep a copy of this notice. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Foundation Medicine apologizes for any inconvenience caused by this issue. If any further information or support concerning this issue is needed, please contact Foundation Medicine at any time.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology