Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1750-2026

Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723) Recall

Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)

FDA recall Class II

Ongoing

FDA lists this Class II recall for Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1750-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 13, 2026
FDA report date
Apr 8, 2026
Quantity in commerce
26,647
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Real-Time Revel Insulin Pump: 00613994951502, 00763000161262, 00613994743930, 00643169077737, 00613994904485, 00643169202399, 00643169151369, 00643169478602, 00643169513914, 00643169933170, 00763000161132, 00643169464520: Artwork, CRB D Process, Paradigm REAL-Time Revel Insulin Pump UG (MP6025306-011A 1); Artwork, CRB D Process, Instr, X23 revel, EN (MP6025306-019 1)

View the original FDA recall notice

Why was this product recalled?

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

What actions did Medtronic MiniMed, Inc. take?

On 2/13/2026, communications began via email, SMS, certified postal mail, and in-app push notifications. Correction notices were sent informing customers of the following: User Guides will be updated to include: "It is important to keep the location of your pump stable relative to your infusion site. Do not wear or place your pump more than 14 in (35.5 cm) above your infusion site. Doing so can cause an over-delivery of insulin, which may result in hypoglycemia." User guides that are updated will be available at: https://www.medtronicdiabetes.com/download-library visit https://info.medtronicdiabetes.com/pump-placement for affected user guide versions. Required Actions: Keep the location of your insulin pump close to your infusion site. - If you need to raise the pump above your infusion site, such as when a parent or caregiver is using the pump for a child, limit the time that the pump is held above the infusion site. - Pay attention to any alerts from your pump as well as symptoms of hypoglycemia and hyperglycemia; follow your healthcare professional's treatment instructions in these situations. - If you have any concerns about how this may impact insulin delivery and blood glucose, please contact your healthcare professional. You may also contact firm at 1-855-587-4950 with any questions about this notification. -Complete and return the acknowledgement form by visiting https://info.medtronicdiabetes.com/pump-placement Distributor Partner / Service Provider: Send existing patients the Patient Notification. To HCPs send the HCP Notification and Patient Notification

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.