Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1754-2025

BIB Balloon in Balloon Dilatation Catheter. Recall

BIB Balloon in Balloon Dilatation Catheter. Model Number: 420

FDA recall Class II

Terminated

FDA lists this Class II recall for BIB Balloon in Balloon Dilatation Catheter.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1754-2025
Recall classification
Class II
Enforcement status
Terminated
Initiation date
Apr 21, 2025
FDA report date
May 21, 2025
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Mixed-up of materials/components

What is being recalled?

BIB Balloon in Balloon Dilatation Catheter. Model Number: 420

Affected model & UDI checker

Search the exact identifiers published for Z-1754-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
420BIB Balloon in Balloon Dilatation Catheter.Check source noticeNot specified
Not specifiedBIB Balloon in Balloon Dilatation Catheter.Check source notice04046964899768
FDA code information

Model Number 420. UDI-DI: 04046964899768. Lot Numbers: BIB-16972 (Expiration date: 11/30/2028), BIB-16972-1 (Expiration date: 12/31/2028). Catheter Product Code: BB005.

View the original FDA recall notice

Why was this product recalled?

Balloon extension lumens are not printed correctly. Specifically, the inner balloon inflation hub was labeled as the outer balloon, and the outer balloon inflation hub was labeled as the inner balloon.

What actions did NuMED, Inc. take?

NuMED sent a "URGENT: MEDICAL DEVICE RECALL" letter dated 4/21/2025 via email on 4/21/2025 to their single customer. The customer was instructed to immediately return the affected units as soon as possible. If you have any questions or concerns, please contact Regulatory Affairs Manager at (315) 328-4491, or via email at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

  5. Recall termination recorded

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology