Bard Single Intragastric Linton Balloon Tube, REF: 0092740 Recall
Bard¿ Single Intragastric Linton Balloon Tube, REF: 0092740
FDA recall Class I
Ongoing
FDA lists this Class I recall for Bard Single Intragastric Linton Balloon Tube, REF: 0092740. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1759-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Apr 17, 2025
- FDA report date
- May 28, 2025
- Quantity in commerce
- N/A
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Bard Single Intragastric Linton Balloon Tube, REF: 0092740
Affected model & UDI checker
Search the exact identifiers published for Z-1759-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 0092740 | Bard Single Intragastric Linton Balloon Tube, REF: 0092740 | Check source notice | Not specified |
| Not specified | Bard Single Intragastric Linton Balloon Tube, REF: 0092740 | Check source notice | 00801741076848 |
FDA code information
All Lots within expiration/UDI: (01)00801741076848
Why was this product recalled?
Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did C.R. Bard, Inc. take?
On April 17, 2025, BD (C.R. Bard) Urology and Critical Care issued a "Urgent Medical Device Correction" Notification to affected consignees. BD asked consignees to take the following actions: 1. Follow the above plastic plug removal instructions. 2. Post this notice where the devices are stored. 3. Please check all inventory locations within your institution for affected products 4. Share this notice with any users of the product within your facilities or with any interfacility users where product was transferred 5.Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of this notification 6.Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088)
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1759-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology