OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem Recall
OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem
FDA recall Class II
Ongoing
FDA lists this Class II recall for OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1766-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 3, 2026
- FDA report date
- Apr 8, 2026
- Quantity in commerce
- 2 kits
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem
Affected model & UDI checker
Search the exact identifiers published for Z-1766-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem | Check source notice | 5032384127498 |
FDA code information
UDI/DI 5032384127498, Lot Number 4494873
Why was this product recalled?
Oxoid Ltd., part of Thermo Fisher Scientific, confirmed that Microbact 12L Kit 20 Tests MB1128A lot 4494873 may identify incorrect organism due to inconclusive result from ambiguous color reactions in wells.
What actions did Oxoid Australia Pty Limited take?
Thermo Fisher Scientific issued an Urgent: Medical Device Recall notice to its consignees on 3/3/2036 via USPS First Class Mail. The notice explained problem, risk to health, requested the consignee destroy any remaining inventory of the affected lot and responding by completing the Acknowledgement form and returning it via email to [email protected] in order to receive replacement/credit to be processed keeping the notification on file. The notice further requests it be passed on to all who need to be aware within the consignee's organization or to any organization where the potentially affected products have been transferred. For questions, please contact our Technical Services Department at 800-255-6730 (US) or email [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1766-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026