Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1767-2026

SIGNA Premier systems Recall

SIGNA Premier systems

FDA recall Class II

Ongoing

FDA lists this Class II recall for SIGNA Premier systems. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1767-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 27, 2026
FDA report date
Apr 8, 2026
Quantity in commerce
41
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

SIGNA Premier systems

Affected model & UDI checker

Search the exact identifiers published for Z-1767-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSIGNA Premier systemsCheck source notice00840682135269
Not specifiedSIGNA Premier systemsCheck source notice00195278010797
FDA code information

UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677

View the original FDA recall notice

Why was this product recalled?

Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Healthcare LLC take?

On 2/27/2026, correction notices were mailed to Director of Clinical/Radiology Risk Manager/Hospital Administrators informing them to take the following actions: 1) Ensure all potential servicing personnel in your facility are made aware of this safety notification and the recommended actions. 2) If the device requires service at the rear of the magnet, ensure that all servicing personnel are aware of this potential issue. 3) If servicing personnel have accessed the rear of the magnet with covers removed and identified that ferrous fittings are installed, do NOT attempt to remove or disconnect them. Call your GE HealthCare representative immediately. To help identify the parts: -The incorrect fittings (ferrous) are steel/silver in color -The correct fittings (non-ferrous) are brass/gold in color 4) Complete and return the acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002kiJR) or print, fill out manually, scan and email to [email protected] If you have any questions or concerns regarding this notification, please contact the firm's HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.