Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252 Recall
Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252
FDA recall Class II
Ongoing
FDA lists this Class II recall for Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1769-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 5, 2026
- FDA report date
- Apr 15, 2026
- Quantity in commerce
- 233
- Distribution
- See key details
- FDA cause
- Labeling design
What is being recalled?
Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
1. UDI-DI (GTIN) 00840338603623 Lot Numbers AZ0124352 AZ0224352 AZ0324352 AZ0424352 AZ0524352 2. UDI-DI (GTIN) 00840338603630 Lot Numbers AZ0624352 AZ0724352 AZ0824352 AZ0924352 AZ1024352 3. UDI-DI (GTIN) 00840338604033 Lot Numbers AZ0124351 AZ0224351 AZ0524351 AZ0724351 AZ0924351
Why was this product recalled?
Affected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are incompatible with the Tornier HRS Max system.
What actions did Tornier, Inc. take?
On March 5, 2026, the firm notified affected customers of the product issue. Customers were instructed to locate affected product in their inventory. Relevant personnel should be informed that the affected product must not be used with Tornier HRS Max products, as they are incompatible. Product return is not necessary, but customers may contact the recalling firm to arrange for return if the product issue impacts their needs.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1769-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology