Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1772-2026

MicroScan Neg MIC 3J REF C54814 Recall

MicroScan Neg MIC 3J REF C54814

FDA recall Class II

Ongoing

FDA lists this Class II recall for MicroScan Neg MIC 3J REF C54814. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1772-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 25, 2025
FDA report date
Apr 15, 2026
Quantity in commerce
688 boxes (20 panels per box)
Distribution
See key details
FDA cause
Process control

What is being recalled?

MicroScan Neg MIC 3J REF C54814

Affected model & UDI checker

Search the exact identifiers published for Z-1772-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C54814MicroScan Neg MIC 3J REF C54814Check source noticeNot specified
FDA code information

Catalog Number: C54814 UDI code: N/A Lot Number: 2025-08-23

View the original FDA recall notice

Why was this product recalled?

Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics

What actions did BECKMAN COULTER, INC. take?

On 07/15/2024, firm sent via mail/email an "URGENT MEDICAL DEVICE RECALL" Letter to customer inform them that customers have reported multiple patient samples with skipped wells for Ceftolozane/Tazobactam (C/T) when tested with MicroScan Neg MIC 3J Lot 2025-08-23. The C/T 2/4 dilution well exhibited no growth while growth was observed in the C/T 4/4 well. Customers are instructed to: "Discontinue use of this lot and discard per your laboratory guidelines. "Consider reviewing the C/T antimicrobial susceptibility test results from this lot if Enterobacterales reported as resistant and P. aeruginosa reported as intermediate or if an increasing trend in resistance has been observed for C/T in your laboratory. "The laboratory should retain the inventory of other MicroScan Neg MIC 3J lots as they are not impacted by this issue. any questions regarding this notice, please contact us: "From our website: http://www.beckmancoulter.com "By phone: please contact: Customer call center TEL :0120-566-730

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology