Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1776-2026

COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791 Recall

COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791

FDA recall Class II

Ongoing

FDA lists this Class II recall for COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1776-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 5, 2026
FDA report date
Apr 15, 2026
Quantity in commerce
49 units
Distribution
See key details
FDA cause
Incorrect or no expiration date

What is being recalled?

COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791

Affected model & UDI checker

Search the exact identifiers published for Z-1776-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedCOOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791Check source notice00827002507876
Not specifiedCOOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791Check source notice00827002507913
FDA code information

Reference Part Number CMW-14-190-12G, UDI (01)00827002507876(17)260509(10)15885574X, Lot Number 15885574X; Reference Part Number CMW-14-190-12G, UDI (01)00827002507876(17)281203(10)15783003X, Lot Number 15783003X; Reference Part Number CMW-14-300-12G, UDI (01)00827002507913(17)260227(10)15416547, Lot Number 15416547.

View the original FDA recall notice

Why was this product recalled?

Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.

What actions did Cook Incorporated take?

Cook Medical issued an URGENT: MEDICAL DEVICE RECALL to its consignee on 3/5 2026 via courier or email. The notice explained the issue, potential risk to health, and requested the following: Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology