FDA RECALL RECORD · Z-1781-2025

Intera CV Recall

Intera CV, Model Number: 781107;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Intera CV. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1781-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 24, 2025
FDA report date
May 21, 2025
Quantity in commerce
6 units (4 US, 2 OUS)
Distribution
See key details
FDA cause
Process control

What is being recalled?

Intera CV, Model Number: 781107;

Affected model & UDI checker

Search the exact identifiers published for Z-1781-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
781107Intera CVCheck source noticeNot specified
FDA code information

Model Number (REF): 781107; UDI-DI: N/A; All serial numbers.

View the original FDA recall notice

Why was this product recalled?

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?

Philips originally notified consignees on 03/24/2025 with Updated URGENT Medical Device Correction letter. It was noted that it contained incorrect information and on 04/09/2025 Philips sent an updated Updated URGENT Medical Device Correction letter. Consignees were instructed to take the following actions prior to a patient scan: 1. Inspect the QBC seal for separation between the cone cover and QBC cover. 2. If QBC seal is found detached or loose, Stop-use immediately. 3. Contact your local Philips service representative. If QBC seal becomes loose during a patient scan: 1. Immediately stop scanning and carefully remove patient from the system. 2. Contact your local Philips service representative. Consignees were also instructed to circulate this URGENT Medical Device Correction letter to all users of this device so that they are aware of the issue, retain this letter with your system(s) until a solution is installed on your system, ensure the letter is in a place likely to be seen/viewed, and complete and return the included response form to Philips promptly and no later than 30 days from receipt via email to: [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.