FDA RECALL RECORD · Z-1805-2025

Product Name: MR 7700; Model Numbers: (1) 782120, (2) 782153; Recall

Product Name: MR 7700; Model Numbers: (1) 782120, (2) 782153;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Product Name: MR 7700; Model Numbers: (1) 782120, (2) 782153;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1805-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 13, 2025
FDA report date
May 28, 2025
Quantity in commerce
72 units (6 US, 66 OUS)
Distribution
See key details
FDA cause
Software design

What is being recalled?

Product Name: MR 7700; Model Numbers: (1) 782120, (2) 782153;

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Model Numbers: (1) 782120, (2) 782153; UDI-DI: (1) 00884838104112, (2) 00884838112858; Serial Numbers: (1) 65000, 65001, 65002, 65003, 65004, 65005, 65007, 65008, 65009, 65011, 65012, 65013, 65014, 65017, 65018, 65019, 65020, 65021, 65023, 65024, 65025, 65026, 65027, 65028, 65029, 65030, 65031, 65032, 65034, 65035, 65037, 65038, 65039, 65040, 65041, 65042, 65050, 65051, 65056, 65057, 65059, 65060, 65061, 65063, 65064, 65065, 65066, (2) 68000, 68001, 68006, 68007, 68008, 68009, 68010, 68011, 68012, 68013, 68014, 68015, 68016, 68017, 68018, 68020, 68023, 68024, 68025, 68028, 68029, 68030, 68031, 68034, 68037;

View the original FDA recall notice

Why was this product recalled?

Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America Llc take?

On May 12, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users You may continue to use your system(s) in accordance with the intended use. A. To avoid the potential issue of an incorrect cross reference line in MobiView generated images, use the same voxel size and field of view across all station scans. B. Do not use cross reference line functionality in box mode, all slices mode, or 3D mode. Use the cross reference line functionality in single slice mode. Circulate this notice to all users of this device so that they are aware of the issues and associated hazard/harm. Please retain this Urgent Medical Device Correction letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed. Please complete and return the attached response form to Philips MR promptly and no later than 30 days from receipt of this letter via email to: [email protected]. Completing this form confirms receipt of the Urgent Medical Device Correction letter, understanding of the issues, and required actions to be taken. 5. Actions planned by Philips MR to correct the problem A Philips representative will contact you to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issues (reference FCO78100566, FCO78100584, FCO78100585, FCO78100620, FCO78100621). If you need any further information or support concerning this issue, please contact your local Philips representative. For North America, contact the Customer Care Solutions Center (1-800-722-9377, 8AM-8PM EST, Monday-Friday). This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax. For more information, pl

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.