TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinician who is responsible for making all patient management decisions. The True 3D Viewer (t3D-Viewer) system is comprised of a commercial off the shelf hardware platform and a proprietary software application that enables a health care professional (HCP) to visualize and interact with CT and MRI DICOM image data to assist in clinical decision making. The t3D-Viewer system hardware platform is comprised of an off the shelf stereoscopic display, an optical or electromagnetic motion tracking system and a computer system. Recall
TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinician who is responsible for making all patient management decisions. The True 3D Viewer (t3D-Viewer) system is comprised of a commercial off the shelf hardware platform and a proprietary software application that enables a health care professional (HCP) to visualize and interact with CT and MRI DICOM image data to assist in clinical decision making. The t3D-Viewer system hardware platform is comprised of an off the shelf stereoscopic display, an optical or electromagnetic motion tracking system and a computer system.
FDA recall Class II
Ongoing
FDA lists this Class II recall for TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinician who is responsible for making all patient management decisions. The True 3D Viewer (t3D-Viewer) system is comprised of a commercial off the shelf hardware platform and a proprietary software application that enables a health care professional (HCP) to visualize and interact with CT and MRI DICOM image data to assist in clinical decision making. The t3D-Viewer system hardware platform is comprised of an off the shelf stereoscopic display, an optical or electromagnetic motion tracking system and a computer system.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1807-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 11, 2016
- FDA report date
- May 28, 2025
- Quantity in commerce
- 9 systems
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinician who is responsible for making all patient management decisions. The True 3D Viewer (t3D-Viewer) system is comprised of a commercial off the shelf hardware platform and a proprietary software application that enables a health care professional (HCP) to visualize and interact with CT and MRI DICOM image data to assist in clinical decision making. The t3D-Viewer system hardware platform is comprised of an off the shelf stereoscopic display, an optical or electromagnetic motion tracking system and a computer system.
Affected model & UDI checker
Search the exact identifiers published for Z-1807-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| EP-00001-XX | TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinician who is responsible for making all patient management decisions. The True 3D Viewer (t3D-Viewer) system is comprised of a commercial off the shelf hardware platform and a proprietary software application that enables a health care professional (HCP) to visualize and interact with CT and MRI DICOM image data to assist in clinical decision making. The t3D-Viewer system hardware platform is comprised of an off the shelf stereoscopic display, an optical or electromagnetic motion tracking system and a computer system. | Check source notice | Not specified |
FDA code information
Model Number EP-00001-XX with software versions 1.0, 1.25, and 1.35 UDI-DI Codes: 00851325007003 Serial Numbers: Field Products: C2015-09-0001 C2015-09-0003 D2015-05-0002 C2015-06-0002 D2015-12-0004 C2015-08-0001 D2015-12-0005 C2015-09-0002 D2015-12-0001 C2016-01-0001 D2015-12-0002 C2015-12-0002 D2015-12-0003 D2016-01-0002
Why was this product recalled?
Imaging system data may be displayed with an incorrect orientation (sidedness) where the patient side is not correct (left, right, superior, inferior, anterior or posterior). This may lead to an inaccurate measurements.
What actions did Echopixel, Inc. take?
On 03/25/2026, the firm sent via email and telephone an Urgent Field Safety Notice, informing customers of a potential misdiagnosis or incorrect surgical treatment planning approach can result when utilizing Enhanced Multi-frame DICOM image data sets. Customers were instructed to: verify if Enhanced Multi-Frame CT DICOM files have been used with the software. If users have utilized this file type, we request that users verify retrospectively any potential risk to health of patients and further determine if this patient needs to be reevaluated and report it to EchoPixel, Inc. We request that the True 3D Viewer versions 1.0, 1.25, 1.35 NOT be used with this file type. For any questions or concerns - contact EchoPixel, Inc Monday thru Friday 9 AM - 6 PM Pacific at: Mailing address: 2490 Hospital Dr. Suite 310, Mountain View CA 94040 Email: [email protected] Phone: +1 (844) 273-7766
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1807-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology