BD Cathena Safety IV Catheter BD Multiguard Technology 18 GA x 1.25 IN, REF: 386865; BD Cathena Safety IV Catheter 20GX1.00IN STRAIGHT BC, REF: 386803; BD Cathena Safety IV Catheter BD Multiguard Technology 20 GA x 1.00 IN, 386862 Recall
BD Cathena Safety IV Catheter BD Multiguard Technology 18 GA x 1.25 IN, REF: 386865; BD Cathena Safety IV Catheter 20GX1.00IN STRAIGHT BC, REF: 386803; BD Cathena Safety IV Catheter BD Multiguard Technology 20 GA x 1.00 IN, 386862
FDA recall Class II
Ongoing
FDA lists this Class II recall for BD Cathena Safety IV Catheter BD Multiguard Technology 18 GA x 1.25 IN, REF: 386865; BD Cathena Safety IV Catheter 20GX1.00IN STRAIGHT BC, REF: 386803; BD Cathena Safety IV Catheter BD Multiguard Technology 20 GA x 1.00 IN, 386862. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1808-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 21, 2025
- FDA report date
- May 28, 2025
- Quantity in commerce
- 262,800
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
BD Cathena Safety IV Catheter BD Multiguard Technology 18 GA x 1.25 IN, REF: 386865; BD Cathena Safety IV Catheter 20GX1.00IN STRAIGHT BC, REF: 386803; BD Cathena Safety IV Catheter BD Multiguard Technology 20 GA x 1.00 IN, 386862
Affected model & UDI checker
Search the exact identifiers published for Z-1808-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 386865 | BD Cathena Safety IV Catheter BD Multiguard Technology 18 GA x 1.25 IN, REF: 386865; BD Cathena Safety IV Catheter 20GX1.00IN STRAIGHT BC, REF: 386803; BD Cathena Safety IV Catheter BD Multiguard Technology 20 GA x 1.00 IN, 386862 | Check source notice | Not specified |
| 386803 | BD Cathena Safety IV Catheter BD Multiguard Technology 18 GA x 1.25 IN, REF: 386865; BD Cathena Safety IV Catheter 20GX1.00IN STRAIGHT BC, REF: 386803; BD Cathena Safety IV Catheter BD Multiguard Technology 20 GA x 1.00 IN, 386862 | Check source notice | Not specified |
FDA code information
REF/UDI-DI/Lot/Expiration: 386865/0100382903868650/4198480/30-Jun-2027, 386803/0100382903868032/4247809/31-Aug-2027, 386862/0100382903868629/4237744/31-Aug-2027
Why was this product recalled?
Due to a manufacturing defect IV catheter may have a hole in the septum, which may result in blood leakage from the septum during insertion, which may cause blood exposure or the need for a second IV to be placed - which may result in therapy delay.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Becton Dickinson Infusion Therapy Systems, Inc. take?
On 4/21/2025, Removal notices were sent (mail/email/fax/phone) to Medical Directors, Risk Managers, Medical Device Safety Officers, Nurse Managers, and distributors were informed of the following: - Routine clinical practice of PPE should protect the clinician from mucocutaneous blood exposure. - If leakage from the septum occurs during insertion, the PIVC should no longer be used. It is recommended the clinician places a new PIVC. In addition, they were asked to do the following: 1) Destroy all unused product subject to the recall following your institution's process for destruction. 2. For indwelling catheters, no further action is needed. 3. If a defective product was previously used on a patient with no issues, no further actions are needed. 4. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 5. Complete and return Customer Response Form via email to Email: [email protected] 6. Distributors were asked to identify all customers within their distribution network that purchased any affected product. Then, Provide a copy of the attached customer letter to all customers to advise them of this field action notification. Questions or complaints can be reported to the North American Regional Complaint Center, 1-844-823-5433 Mon Fri 8:00am and 5:00pm CT or Email: [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1808-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology