iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software Recall

iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software

FDA recall Class II

Ongoing

FDA lists this Class II recall for iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1809-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 13, 2025
FDA report date
Apr 22, 2026
Quantity in commerce
15 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software

Affected model & UDI checker

Search the exact identifiers published for Z-1809-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
BB1001iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision SoftwareCheck source noticeNot specified
FDA code information

ILet Ace Pumps containing software version BOM000036 Rev C (version 1.4.2) or older.

View the original FDA recall notice

Why was this product recalled?

it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.

What actions did Beta Bionics, Inc. take?

On March 12, 2026 Beta Bionics issued a Urgent Medical Device Correction - Expansion/Update via Email to notify new consignees and previous consignees. Initial notification was issued on August 13, 2025. Beta Bionics ask consignees to take the following actions: 1. Download or update your iLet app on your phone. Read the iLet App User Guide. 2. Log into your iLet app. 3. Position your iLet next to your phone with the app open. 4. Go to app settings on the iLet app and follow the on-screen instructions to update to the latest firmware. 5. Complete the acknowledgment survey https://sprw.io/stt-6HnQU or scan the QR code. This survey helps us monitor the effectiveness of the recall action.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology