ARTIS Pheno VE30A and VE40A, Model 10849000 Recall

ARTIS Pheno VE30A and VE40A, Model 10849000

FDA recall Class II

Ongoing

FDA lists this Class II recall for ARTIS Pheno VE30A and VE40A, Model 10849000. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1814-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 12, 2026
FDA report date
Apr 22, 2026
Quantity in commerce
73
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

ARTIS Pheno VE30A and VE40A, Model 10849000

Affected model & UDI checker

Search the exact identifiers published for Z-1814-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
10849000ARTIS Pheno VE30A and VE40A, Model 10849000Check source noticeNot specified
Not specifiedARTIS Pheno VE30A and VE40A, Model 10849000Check source notice04056869046877
FDA code information

UDI 04056869046877

View the original FDA recall notice

Why was this product recalled?

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Medical Solutions USA, Inc. take?

Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers dated March 9, 2026.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.