Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1828-2026

CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system. Recall

CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.

FDA recall Class II

Ongoing

FDA lists this Class II recall for CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1828-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 6, 2026
FDA report date
Apr 29, 2026
Quantity in commerce
78 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.

Affected model & UDI checker

Search the exact identifiers published for Z-1828-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C7000CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.Check source notice10381780126232
FDA code information

Model: C7000; UDI: 10381780126232; Serial Numbers: CHA1902101IE, CHA2001301IE, CHA2001401IE, CHA2002301IE, CHA2202301IE, CHA2305101IE, CHA2305401IE, CHA2307001IE, CHA2402201IE, CHB1711001IE, CHB1723601IE, CHB2000801IE, CHB2100901IE, CHB2301401IE, CHB2500101IE, CHC1800601IE, CHC1902601IE, CHC1902701IE, CHC2001201IE, CHD1705801IE, CHD1707101IE, CHD1801001IE, CHD1801101IE, CHD1902501IE , CHE1707201IE, CHE1707301IE, CHE2203301IE, CHE2204001IE, CHE2301001IE , CHE2305001IE, CHE2502901IE, CHE2503101IE, CHE2503201IE, CHE2503401IE, CHE2503701IE, CHF1701401IE, CHF1702701IE, CHF2100701IE, CHF2303701IE, CHF2305401IE, CHG1700101IE, CHG1700801IE, CHG1902401IE, CHG2000201IE, CHG2000301IE, CHG2204101IE, CHG2402301IE, CHG2502801IE, CHG2506001IE, CHH1802901IE, CHH2402801IE, CHH2500601IE, CHJ2001101IE, CHJ2001301IE , CHJ2301201IE, CHJ2304601IE, CHK1700801IE, CHK1702301IE, CHK1703001IE, CHK2402601IE, CHK2403401IE, CHK2404101IE, CHL1701701IE, CHL1900201IE, CHL2102901IE , CHL2103001IE, CHL2301301IE, CHL2302001IE, CHL2302101IE, CHL2403301IE, CHM1901901IE, CHM2100101IE, CHM2101601IE, CHM2200201IE, CHM2200301IE, CHM2200501IE, CHM2300701IE, CHM2402101IE.

View the original FDA recall notice

Why was this product recalled?

Software issue that renders the touch screen unresponsive.

What actions did Integra LifeSciences Corp. (NeuroSciences) take?

An URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice dated 3/6/26 was sent to consignees. The notice instructs consignees to restart systems when freeze to restore functionality of the touchscreen capabilities for 8 minutes; this action may be used multiple times. Integra's Service and Repair can provide remote support to disable and re-enable the write filter with Integra employees and/or Authorized service personnel. This will reset the console to provide approximately 28 days with no touchscreen issue. Integra Lifesciences Service Team will contact consignees within 5 to 10 business days to schedule correction of affected devices at no cost. Consignees with any questions can direct them to Customer Service by emailing [email protected] or by phone at 1-800-654-2873.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology