Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm. Recall
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm. Model Number: TB2-0535FC.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1853-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 23, 2026
- FDA report date
- Apr 29, 2026
- Quantity in commerce
- 3,360 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm. Model Number: TB2-0535FC.
Affected model & UDI checker
Search the exact identifiers published for Z-1853-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| TB2-0535FC | Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm. | Check source notice | Not specified |
| Not specified | Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm. | Check source notice | 04953170440014 |
FDA code information
Model Number: TB2-0535FC. UDI-DI: 04953170440014. All Lot Numbers.
Why was this product recalled?
Potential for detachment of a distal tip component of the device during use.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Olympus Corporation of the Americas take?
Olympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were transferred or further distributed.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1853-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026