Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1855-2026

Intuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06 Recall

Intuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06

FDA recall Class I

Ongoing

FDA lists this Class I recall for Intuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1855-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Mar 11, 2026
FDA report date
May 13, 2026
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Intuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06

Affected model & UDI checker

Search the exact identifiers published for Z-1855-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06Check source notice00886874121931
FDA code information

UDI:00886874121931/All serial Numbers

View the original FDA recall notice

Why was this product recalled?

Reloads for curved-tipped stapler may potentially produce an incomplete staple line.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Intuitive Surgical, Inc. take?

On 3/17/2026 Intuitive Surgical issued a "urgent Medical Device Recall Notification to affected consignees via E.Mail and FedEx. Intuitive ask consignees to take the following actions: 1. Please identify, discontinue use, and quarantine any affected product(s). Affected product(s) may be returned by calling customer service with quantities and lot numbers to your regional customer service. 2. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 3. Please ensure to include the Recall or FSCA number ISIFA2026-02-R in your return notes. 4. If you have shared or further distributed these products with other sites, please make sure appropriate staff at the site receive and understand this notification so they locate and return their affected product.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.