Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system Recall
Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system
FDA recall Class II
Ongoing
FDA lists this Class II recall for Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1859-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 31, 2025
- FDA report date
- Jun 4, 2025
- Quantity in commerce
- 96 systems
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system
Affected model & UDI checker
Search the exact identifiers published for Z-1859-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 6003000804 | Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system | Check source notice | Not specified |
| 6003000805 | Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system | Check source notice | Not specified |
| Not specified | Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system | Check source notice | 00816634020013 |
| Not specified | Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system | Check source notice | 00816634020006 |
FDA code information
Triton 36 Model Number 6003000804 UDI-DI code: 00816634020013 Serial Numbers: 1601-XL36-249-1 1604-XL36-277-1 1808-XL36TR-596-1 1810-XL36TR-618-1 1905-XL36TR-707-1 1907-XL36TR-721-1 1911-XL36TR-800-1 2001-XL36TR-834-1 2004-XL36TR-859-1 2004-XL36TR-873-1 2104-XL36TR-947-1 2201-XL36TR-001-1 2203-XL36TR-017-1 2306-XL36TR-050-1 2306-XL36TR-051-1 2307-XL36TR-053-1 2312-XL36TR-120-1 2401-XL36TR-001-1 2402-XL36TR-017-1 2402-XL36TR-018-1 2402-XL36TR-027-1 2402-XL36TR-028-1 2402-XL36TR-029-1 2402-XL36TR-030-1 2405-XL36TR-048-1 2405-XL36TR-049-1 2405-XL36TR-050-1 2405-XL36TR-051-1 2405-XL36TR-060-1 2405-XL36TR-061-1 2405-XL36TR-062-1 2405-XL36TR-063-1 2406-XL36TR-064-1 2406-XL36TR-065-1 2406-XL36TR-072-1 2406-XL36TR-074-1 2406-XL36TR-075-1 2407-XL36TR-079-1 2407-XL36TR-083-1 2409-XL36TR-101-1 2409-XL36TR-103-1 2409-XL36TR-105-1 2409-XL36TR-107-1 2409-XL36TR-109-1 2409-XL36TR-110-1 2410-XL36TR-111-1 2410-XL36TR-113-1 2410-XL36TR-114-1 2411-XL36TR-118-1 2411-XL36TR-122-1 2412-XL36TR-135-1 XL36TR4381 XL36TR4961 XL36TR5111 2307-XL36TR-055-1 Triton 72 Model Number 6003000805 UDI-DI code: 00816634020006 Serial Numbers: 1511-XL72-230-1 1606-XL72-291-1 1706-XL72TR-401-1 1805-XL72TR-544-1 1805-XL72TR-546-1 1806-XL72TR-564-1 1806-XL72TR-565-1 1807-XL72TR-568-1 1809-XL72TR-611-1 1910-XL72TR-788-1 2002-XL72TR-844-1 2004-XL72TR-878-1 2004-XL72TR-879-1 2005-XL72TR-881-1 2006-XL72TR-890-1 2103-XL72TR-929-1 2103-XL72TR-930-1 2104-XL72TR-946-1 2112-XL72TR-1022-1 2202-XL72TR-013-1 2202-XL72TR-014-1 2202-XL72TR-015-1 2203-XL72TR-016-1 2207-XL72TR-045-1 2211-XL72TR-088-1 2304-XL72TR-028-1 2305-XL72TR-030-1 2305-XL72TR-034-1 2309-XL72TR-073-1 2310-XL72TR-099-1 2402-XL72TR-023-1 2402-XL72TR-024-1 2404-XL72TR-047-1 2405-XL72TR-052-1 2405-XL72TR-053-1 2406-XL72TR-073-1 2407-XL72TR-080-1 2407-XL72TR-084-1 2409-XL72TR-104-1 2411-XL72TR-120-1 XL72TR4401
Why was this product recalled?
During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.
What actions did Ultra Clean Systems, Inc. take?
On 03/31/2025, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx to customers informing them that as a result of an investigation, it has been revealed that a combination of chemical and electrical interactions that result in electrochemical degradation, compromising the structural integrity of the sensor over time. This degradation could potentially lead to water leakage into the machine. On 04/25/2025, the firm sent via FedEx and updated letter. Customers are notified and/or instructed that/to: -A field technicians will be contacting you to arrange for replacement if your unit has been impacted by this issue. -Complete and sign the attached response form to acknowledge that they have received the notification. Return the completed form to Getinge by emailing a scanned copy to [email protected] Public contact for questions/assistance - [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1859-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology